目的:探讨胸部摄影AEC自动曝光控制下影像密度与焦-片距的相关性。方法:利用胸部摄影自动曝光控制装置,使用不同的焦-片距离,摄取正常的标准胸片24张,确定7个评价测试点,用密度计分别测量7个测试点的密度值。结果:肺野的密度与焦-片距离没有密切的相关性(r<0.811P>0.05只有气管分叉处的密度值与焦-片距之间有非常显著的负相关(r=-0.955P<0.01)。结论:考虑病人的吸收剂量和避免影像失真,摄影焦-片距的选择应与滤线栅的焦距相匹配。
Abstract:Objective: To investigate the correlation of image density with focus-film distance in automatic exposure control(AEC) of chest radiography. Method: Taking the focus-film distance as a single varian, the 24 chest radiographs in various focus-film distance were selected according to AEC standard and we confirmed 7 evaluation points were selected and the density of each evaluation point were measured by densitometer. Results: It showed that the density of lung field was not closely correlated with the focus-film distance (all r<0.811, all P >0.05) except at the area of trachea branch (r<-0.955,P <0.01). Conclusion: Considering the absorbed dose of the patients and avoiding the image distortion, It is suggested that the selection of focus-film distance should be matched with the tube and grid.
1郁鹏,李伟.胸部高电压摄影自动曝光控制下探测点选择方式的探讨.中华放射学杂志,1999,33(5):344~345
2朱宗平.自动曝光控制成像因素的应用分析.中华放射学杂志,199024(5286~288
3EISMAILOSIMASTORAKOUNLKELEKIS.etal.Clinicalevaluationofmanualandautomaticexposurecontroltechniquesinfilm-basedchestradiography.BJR199669650~654
4尉可道.AEC对放射诊断摄影影像的质量评价研究进展.中华放射学杂志,1997,31(1):10~13
5AJHUNTSGPLAIN.Asimplesolutiontotheproblemsoftestingautomaticexposurecontrolindiagnosticradiology.BJR199366360~362
6袁聿德.X线摄影学.第2版.北京:人民卫生出版社,1999.83~84
7王鹏程,刘世民,王洪,等.X线自动曝光控制电离室的研制及其特性研究.中国医疗器械杂志,1999,23(1:30~34
8钟敦荫.现代X线机电路分析与维修(岛津机型选).武汉武汉工业大学出版社,1992.383~386
基本信息:
DOI:10.13210/j.cnki.jhmu.2001.03.008
中图分类号:R814.3
引用信息:
[1]赵军,林卫,吴礼胜,杨进全,龙甫桀.胸部摄影自动曝光控制装置影像密度与焦-片距的相关性研究[J].海南医学院学报,2001(03):147-149.DOI:10.13210/j.cnki.jhmu.2001.03.008 .
2001-09-30
2001-09-30